CSlaser 2M

IR thermometer optris CSlaser 2M for measurement of metallic surfaces

The infrared thermometer optris CSlaser 2M has been designed specifically for exact temperature measurement of metallic surfaces. Its short measuring wavelength allows not only for precise measurement of metal temperatures but metal oxides and ceramics as well.

The IR thermometer is equipped with an innovative double laser sight for precise marking of the measuring spot. A variety of scopes ensures high adaptability with various applications.

Important specifications

  • Temperature range:
    250°C to 1,600°C
  • Spectral range: 1.6 µm
  • Response time: 10 ms
  • Starting at 560 €

From 610,00 

Technical Details of the optris CSlaser 2M

Measurement Specifications

  • Temperature range (scalable via software):
    250°C … 800°C (2ML)
    385°C … 1600°C (2MH)
  • Spectral range: 1.6 µm
  • Optical resolution (90% energy):
    150:1 (2ML)
    300:1 (2MH)
  • System accuracy (at Tamb = 23 ± 5°C):
    ± (0,3 % of reading + 2 °C)
  • Repeatability (at Tamb = 23 ± 5°C):
    ± (0.1 % of reading + 1 °C)
  • Temperature resolution: 0.1 K
  • Response time (90% signal): 10 ms
  • Emissivity / Gain (adjustable via sensor or software): 0.100 – 1.100
  • IR window correction (adjustable via software): 0.100 – 1.000
  • Signal processing (parameter adjustable via software):
    peak hold, valley hold, average, extended hold function with threshold and hysteresis

 

General Specifications

  • Environmental rating: IP 65 (NEMA-4) front mountable at vacuum processes (up to 10-3 mbar)
  • Ambient temperature: -20°C … 85°C (50°C with laser ON)
  • Storage temperature: -40°C … 85°C
  • Relative humidity: 10 – 95%, non condensing
  • Vibration: IEC 60068-2-6 (sinus shaped), IEC 60068-2-64 (broadband noise)
  • Shock: IEC 60068-2-27 (25G and 50G)
  • Weight: 600 g

 

Electrical Specifications

  • Output / analog: 4-20 mA
  • Loop resistance: max. 1000 Ω1)
  • Output / alarm: 0-30 V/500 mA (open collector)
  • Output/ digital: uni- / bidirectional, 9.6 kBaud, 0/3 V digital level / USB optional
  • Cable length: 3 m, 8 m, 15 m
  • Power supply: 5-30 V DC
  • Current draw:
    45 mA at 5 V
    20 mA at 12 V
    12 mA at 24 V

 

Versions

Part number: temperature range / spectral range / optics

  • OPTCSL2ML: 250-800°C / 1.6 μm / 150:1
  • OPTCSL2MH: 385-1,600°C / 1.6 μm / 300:1

 

1) in dependence on supply voltage

Areas of application for the infrared thermometer optris CSlaser 2M

The optris CSlaser 2M offers a short-wave spectral range of 1.6 µm and is thus specifically suited for temperature measurements of metal surfaces, metal oxides and ceramics.

Infrared software optris Compact Connect included

The temperature analysis software optris Compact Connect has been developed specifically for the extensive documentation and analysis of temperature measurement data of our stationary infrared thermometers. It is license-free and included in every pyrometer shipment.

Accessories